Digital Display
Four Digit (LED), with 0.2 Second Update
Dimensional Readout
English and Metric (Selector Switch)
Gauging Range
English: 0.005-10 inches Metric 0. 13-100 mm (Depending Upon Material)
Digital Resolution
English Units
±0.0001 " on 1 " Range
±0.001 " on 10" Range
Metric Units
±0.0001mm on 10 mm Range
±0.001mm on 100 mm Range
Measuring Mode
Sets Mode for Contact, Delay Line or Immersion Transducers
Readout Mode
Selects Digital Display for Thickness or Velocity
Pulser
Risetime
5 to 10 ns into 50 Ohms (Depending on Pulser Voltage)
Amplitude
Selectable for 90, 150, or 300 Volts
Peak into 50 Ohms
Repetition Rate
Selectable for 625, 1200, 2500 or
5000 Hz
Damping
15 to 350 Ohms
Receiver
Adjustable Gain
66dB
Automatic Gain Control
40 dB Dynamic Range
Reject
Variable Threshold to Full Scale
Bandwidth
30 MHZ (6 dB down),
20 MHZ (3dB down)
Attenuator
0, 10, and 20 dB Selectable
A-Trace Display
Sweep Width
Switchable down to 50 ns/Div Continuously Adjustable
Sweep Delay
- 1 ms to 80 ms
Display
- Dual Trace; A-Trace plus Selectable
Gate Traces (IP, IF, T or TAC)
Display Modes
- Full Wave RF, Positive or Negative
Rectified
T-Gate
Thickness Gate with Adjustable Sync
T-Gate Start/Stop
Selectable on Positive or Negative Half Cycles
IP-Gate
0.25 to 20 ms (Contact), 1.0 to 0 ms (Delay or Immersion)
IF Gate
0.1 to 8.0 ms
TAC-Gate
Adjustable thickness Amplitude Compensation of
Gain:
Start Control0.02 to 6 ms
Amplitude Control 0 to 17 dB
Slope Control0.1 to 50 ms
Alarms
Three functions: LO, HI, LO/Hl
Outputs
Alarms, Analog Thickness, Digital High Speed Binary, Serial RS-232C, IP Sync and PortaScan
Inputs
External Sync
Power Source
Switchable Between Nominal 115VAC (95-128) and Nominal 230VAC (185-250) 48 to 440 Hz
Power Consumption
70 Watts
Size
5.4"H x 12.9" W x 17.2 D
(137 x 328 x 437mm)
Weight
18.2 lbs. (8.3 kg)
Warranty
One Year
Bubbler Series
NDT Systems manufactures a vast array of bubbler style thickness gaging transducers. These transducers are optimized for the NovaScope generation of thickness gages but work quite well with virtually any A-Trace based thickness gages.
Water Recirculation System
T300 Water Recirculation System. Generally used with bubbler style transducers (one is shown in the center top... optional). The T300 permits a gentle water flow over the face and through a water column of captive water column transducers such as the IBU25. Water then flows back into the tank where it is collected for reflow.
|
|
|
|
|
|
NovaScope 5000